.

 

Home
310-348-9656
Contact Us
Get a Quote
Give us Your Challenge
Product Overview
Standard Products
BM-10 Platform-Bench Top
BM-8 Platform-Bench Top
BM-6 Platform-Bench Top
BM-4 Platform-Bench Top
BM-1 Platform-Bench Top
BA-1  Platform-Bench Top
MK26 Table-Workstation
MK52 Optical Table
WS4 Table-Workstation
CM-1 Compact
CT-1 Ultra-Thin
LC-4 Ultra Compact
SM-1 Large Capacity
FP-1 Floor Platform
Manuals & Documents
What's the Right System
Custom Systems
Applications
Microscopy
Micro-Hardness Testing
Optical & Laser Systems
Spacecraft Testing
Biology & Neuroscience
Microelectronics & MEMS
Analytical Balances
Audio Reproduction
Vacuum Isolation
Large-Displacement
Heavy Systems
About Us
Our Technology
FAQs
Press & News
Case Studies
Performance
Customers
Testimonials
Glossary
Videos
Educational Giveaway
Newsletters
Sitemap

Newsletter November 2019
"VC-M is the lowest we have ever measured, though we weren’t able to measure below 2.5 Hz because our most sensitive sensor won’t go lower, due to sensor noise floor."  more...

- Vibration Criterion (VC) Curves-Lab Analysis

- Featured Product: WS-4 Compact Vibration Isolation Table
- FAQs>>Frequently Asked Questions About Vibration Isolation

- How They Work>>Negative-Stiffness Vibration Isolators

- Custom Vibration Isolation Systems

- Audio Reproduction & Turntable Vibration Isolation

- Microscopy Vibration Isolation

- 300 leading universities and private and government laboratories
in 51 countries use Minus K technology


Give us your Challenge                Pricing               Get a Quote




 

Vibration Criterion (VC) Curves-Lab Analysis
Codes and curve descriptions for different vibration environments
and enviromental solutions.

The VC (Vibration Criteria) curves were developed in the early 1980s by Eric Ungar and Colin Gordon. They were originally developed as a generic vibration criteria for vibration-sensitive equipment for use in the semiconductor, medical and biopharmaceutical industries, but have found application in a wide variety of technological applications.

The criteria takes the form of a set of one-third octave band velocity spectra, together with the International Standards Organization (ISO) guidelines for the effects of vibration on people in buildings. The criteria apply to vibration as measured in the vertical and two horizontal directions.

The NIST-A criterion was developed for metrology, but has gained popularity within the nanotechnology community. The NIST-A criterion is a very difficult criterion to meet at some sites with significant low-frequency vibrations.

The VC curves are now widely accepted throughout the world as a basis for designing a facility to meet the requirements of a group of highly vibration sensitive equipment used close together.

University of Michigans Ultra-Low Vibration Lab (ULVL) was completed in 2014. After the construction, a vibration survey was done on the Ultra-Low Vibration Lab chambers. The measurements demonstrated that even when a single vehicle was driving on a nearby street, the vibrations exceeded the NIST-A specifications necessary for the ULVL.

The University of Michigan ordered seven customized tabletops and 31 custom Minus K Negative-Stiffness vibration isolators with pedestals provided for the eight Ultra-Low Vibration Lab chambers.


Customized Minus K Technology Negative-Stiffness vibration isolation table installed in one of the Ultra-Low Vibration Lab chambers

The final vibration survey by Colin Gordon Associates (CGA), after installation of the customized Minus K Negative-Stiffness isolators and tables, showed the measured vibration levels in all ULVL chambers from VC-K to VC-M at frequencies above 2.5 Hz, well below the NIST-A Vibration Criterion required.

"VC-M is the lowest we have ever measured, though we werent able to measure below 2.5 Hz because our most sensitive sensor wont go lower, due to sensor noise floor," said Hal Amick, Vice President of Colin Gordon Associates.

The updated VC Curve on Minus K's website shows these lower curve levels that were measured by CGA and have already assisted University of Michigans ULVL with two major scientific milestones.


Vibration site surveys can tell you a lot about how to
specify equipment for vibration isolation in your laboratory.

Curve Criterion Amplitude1
µm/s (µin/s)

Detail Size2
µm

Description of Use
Workshop (ISO)
800( 32,000)
N/A
Distinctly perceptible vibration. Appropriate to workshops and non-sensitive areas.
Office (ISO)
400 (16,000)
N/A
Perceptible vibration. Appropriate to offices and non-sensitive areas
Residential Day (ISO)
200 (8000)
75
Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment and microscopes less than 40x.
Op. Theatre (ISO)
100 (4000)
25
Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100X and for other equipment of low sensitivity.
VC-A
50 (2000)
8
Adequate in most instances for optical microscopes to 400X, microbalances, optical balances, proximity and projection aligners, etc.
VC-B
25 (1000)
3
Appropriate for inspection and lithography (including steppers) to 3 µm line widths..
VC-C
12.5 (500)
1 - 3
Appropriate standard for optical microscopes to 1000X, inspection and lithography inspection equipment (including moderately sensitive electron microscopes) to 1 µm detail size, TFT-LCD stepper/scanner processes.
VC-D
6.25 (250)
0.1 - 0.3
Suitable in most instances for the most demanding equipment including electron microscopes (TEMs and SEMs) and E-Beam systems..
VC-E
3.12 (125)
< 0.1
A challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability.
VC-F
1.56 (62.5)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-G
.78 (31.3)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
NIST-A
NIST-A criterion is identical to the VC-E curve at frequencies above 20 Hz, but maintains constant displacement at frequencies below this. 0.025 µm or 25 nm between 1 and 20 Hz; 3.1 µm /s (125 µ in /s) between 20 and 100 Hz
NIST-A1
NIST-A1 criterion requires an RMS velocity of 3 µm/sec (118 µin./sec) for frequencies less than 4 Hz, and 0.75 µm/sec (29.5 µin./sec) velocity for frequencies 4 Hz < f < 100 Hz
VC-H
.39 (15.63)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-I
.195 (7.81)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-J
.097 (3.9)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-K
.048 (1.95)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-L
.024 (0.98)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
VC-M
.012 (0.49)
N/A
Appropriate for extremely quiet research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation.
1 As measured in one-third octave bands of frequency over the frequency range 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G)
2 The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc. It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.
The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.
This table is a reference of Colin Gordon Associates ,except (3&4).
3 Vibro-Acoustic Consultants, 4 Purdue University Libraries.
.
5 NIST-A1 criterion was for the NIST AML (0.75 mm/s at frequencies between 4 and 100 Hz). NIST AML has a quantity of vibration-isolated spaces, created by placing large concrete blocks on air springs. If these had been used in a space just meeting NIST-A, then the performance would be on the order of NIST-A1, accounting for degradation at internal resonances. This was more of a philosophy than a set criterion amplitude.

Updated VC Curve on Minus K's website...


Give us your Challenge               Pricing               Get a Quote

 

Featured Product: WS-4 Compact Vibration Isolation Table

  • Vertical natural frequency of 1/2 Hz or less can be achieved over the entire load range.
  • Horizontal natural frequency is load dependent. 1/2 Hz or less can be achieved at or near the nominal load.
  • Horizontal natural frequency is load dependent. 1/2 Hz or less can be achieved at or near the nominal load.
  • Dimensions: 24" W x 24" D x 23" H
    (610mm W x 610mm D x 533mm H)

A cost effective vibration isolation table for weight loads up to 1000 lb. and 1/2 Hz performance vertical and horizontal.

The stand has been designed with a damped Phenolic top which includes locating holes for the leveling feet of the BM-1 and BM-4.

Pricing & Specifications



Give us your Challenge               Pricing               Get a Quote
 

FAQs>>Frequently Asked Questions About Vibration Isolation

       What question do "you" have for us?


Give us your Challenge               Pricing               Get a Quote

 

How They Work>>Negative-Stiffness Vibration Isolators

Minus K® vibration isolators employ a revolutionary concept in low-frequency vibration isolation. Vertical-motion isolation is provided by a stiff spring that supports a weight load, combined with a negative-stiffness mechanism (NSM). The net vertical stiffness is made very low without affecting the static load-supporting capability of the spring. Beam-columns connected in series with the vertical-motion isolator provide horizontal-motion isolation. The horizontal stiffness of the beam-columns is reduced by the "beam-column" effect. (A beam-column behaves as a spring combined with an NSM.) The result is a compact passive isolator capable of very low vertical and horizontal natural frequencies and very high internal structural frequencies.

Figure 1

  Minus K® isolators typically use three isolators stacked in series:
A tilt-motion isolator on top of a horizontal-motion isolator on top of a vertical-motion isolator. A vertical-motion isolator is shown in
Figure 1. It uses a conventional spring connected to an NSM consisting of two flexures connected at their inner ends to the spring and supported at their outer ends, and loaded in compression by forces P. The spring is compressed by weight W to the operating position of the isolator, as shown in Figure 1. The stiffness of the isolator is K=KS-KN where KS is the spring stiffness and KN is the magnitude of a negative stiffness which is a function of the design of the flexures and the load P. The isolator stiffness can be made to approach zero while the spring supports the weight W.

A horizontal-motion isolation system consisting of two beam-column isolators is shown in Figure 2. Each isolator behaves like two fixed-free beam columns loaded axially by a weight load W. Without the weight load the beam-columns have horizontal stiffness KS. With the weight load the lateral bending stiffness is reduced by the "beam-column" effect. This behavior is equivalent to a horizontal spring combined with an NSM so that the horizontal stiffness is K=KS-KN, and KN is the magnitude of the beam-column effect. Horizontal stiffness can be made to approach zero by loading the beam-columns to approach their critical buckling load.

 
Figure 2

Full article + more images...

Performance...


Give us your Challenge               Pricing               Get a Quote

 
Custom Vibration Isolation Systems

Minus K engineers can design a variety of custom vibration isolation systems. Frequently these are modifications of other designs such as changing the load capacity of an SM-1 isolator, or the capacity, size and isolator locations on a FP-1.


Custom table top with four resessed Minus K CM-1s and pedestals

Some other custom modifications and capabilities include:

  • Different vertical, horizontal or tilt stiffness in an isolator
  • Different vertical, horizontal or tilt damping in an isolator
  • Vacuum compatibility
  • Mildly magnetic
  • Heavy load capacities (Six 10,000 lb isolators for NASA's James Webb
    Space Telescope Ground Testing
    ).
  • Large displacements
  • Large temperature variations
  • Compact systems (e.g., a seven-inch cubic 1/2 Hz isolator for a 700 lb payload;
    a six-inch diameter by 4-inches tall 1/2 Hz vibration isolation platform for a 3 lb payload.)
  • Vertical auto-adjust feature to automatically adjust for change in payload weight.
  • Lock-unlock feature implemented manually with a switch or automatically
    with signals, from a stage motion control system, for example.

Give us your Challenge               Pricing               Get a Quote

 
Audio Reproduction & Turntable Vibration Isolation
Platforms, Tables, & Integrated Systems


Since the early 2000s, Minus K Technology, the developer of Negative-Stiffness vibration isolation, has released several vibration isolation platforms for the high-end audio market. These products have been accepted with high acclaim by audiophiles and audio system manufacturers worldwide.

The industry's highest performing vibration isolation system is available in a cost-effective, low-height platform. The BM-8 Series vibration isolation platform offers true ½-Hz isolation performance vertical and 1.5 Hz horizontal and is easy to use. No air or electric power is required, there is no maintenance and nothing to wear out. It is 18Wx20Dx4.6H and is available with nominal load capacities from 25 to 250 lbs. It is ideal for turntables, CD transports and other audio components.

The Helix One® and Helix Two® turntables from Dohmann Audio, for example, come fully integrated with Negative-Stiffness vibration isolation.

Doehmann Helix 1 (BM-8 fully integrated)

Doehmann Helix 2 (BM-10 fully integrated)

“The ability of Negative-Stiffness to deal with vibration and resonance is unsurpassed, and allows the music to be enjoyed with an outstanding level of realism.

Analog audiophile aficionados talk about blackberry notes, where the base is too prominent and muddy, explained Dohmann. This is often caused by a reinjection of vibration into the playback mechanism, adding to the amplitude of the needle movement and creating more signal that what is physically in the groove. But Negative-Stiffness vibration isolation cleans this up, so that the base is more natural. The timbre of the instrument used to produce the base, whether a string, piano or drum, sounds more like the real thing. It enables a sense of peace and calm to the playback."

  - Mark Döhmann, Founder Döhmann Audio

The latest generation of Negative-Stiffness vibration isolation technology, embodied within the new ultra-thin, low-hieght Minus K CT-1, truly represents a significant advance in sound quality in a very compact package for high-end audio perfectionists.

SAT (Swedish Analog Technologies) XD1 Record Player
System with Minus K CT-1 [integrated into the platform]


“I have now tried the CT-1 with the SAT top platform the results are spectacular! I took a look at the mechanism inside of the CT-1 and I have to say I am very impressed with the ingenious work you have done. It provides most of its big brothers performance in an extremely compact package that looks very slick. The arrangement and the finish of the mechanism is an engineers dream and it should be proudly displayed. The equipment supported by the CT-1 is well isolated from vibrations from 0.7 to 100 Hz, with much deeper, clearer base and more information in the lower frequencies. The CT-1 makes a huge difference in sound quality."
  - Marc Gomez, President / Swedish Analog Technologies SAT
The BM-1 Series platform offers Minus K's signature 0.5 Hz performance, both vertical and horizontal, and is available for larger and much heavier audio components.

The BM-10 Series offers smaller horzontal dementions at 12.2 W x 12.2 D x 4.6 H (310mm W x 310mm D x 117mm H).

More images, plus reviews and testimonials...



Give us your Challenge               Pricing               Get a Quote

 
Microscopy Vibration Isolation
Tables, Platforms, Benchtops & Custom Systems for
SEM, AFM, TEM, SPM, STM, STEM, NSOM-SNOM



Zeiss Axio Obersver Z1 AFM
on a Minus K BM-8


Delong LVEM5 Desktop SEM
on a Minus K BM-8


Nion STEM on a four-isolator
Minus K FP-1


NanoMagnetics ezAFM
on a Minus K 50BM-4


Many of our clients' microscopes have nano and micro vibration isolation requirements that are unparalleled in the metrology world (for brain research and at NASA as examples). Minus K's negative stiffness and "passive" verses "active" technology, enables microscopes to achieve their highest level of performance. Our isolators and custom systems provide the performance of an ideal bungee system and the convenience of a bench top system without any of the problems associated with many of them, including air tables.

From the ubiquitous bungee suspension systems, to large beds of sand on inner tubes, there have been many attempts at achieving the ultimate isolation. Air tables have been used with some success. However, our negative-stiffness isolators provide superior isolation control and performance while offering better ease of use and no facility requirements. We even have the flexibility of custom tailoring resonant frequencies vertically, horizontally and in tilt to your liking.

Small "active" systems (as opposed to our "passive" systems) are sometimes used. However, those have several drawbacks:
  • They are expensive and fragile.
  • They are subject to the noise floor of every component in the entire feedback loop plus AC line noise.
  • They have a limited dynamic range, typically only a few thousandths of an inch. If this is exceeded (very easy to do) the isolator goes into positive feedback and you have a very expensive noise generator underneath your SPM.
  • They have to be sitting on an infinitely rigid structure to work at their theoretical best. Obviously, infinitely rigid structures are not easy to find in most labs. The result is that the isolator is not always doing what you think it's doing.
Our isolators give you the best possible performance in all axes, from high to low frequency isolation. They provide the performance of an ideal bungee system and the convenience of a bench top system without any of the problems associated with any of them, including air tables. We enable our users to achieve the highest level of noise performance possible.


More images...

Microscope isolation testimonials...



Minus K Customers:
With users at more than 300 leading universities and private and government laboratories in 51 countries, Vibration Isolation News is designed to keep our customers and friends up to date on the latest products and applications designed to facilitate better measurements and improved nanomanufacturing. We are an OEM supplier to leading manufacturers of scanning probe microscopes, micro-hardness testers and other sensitive instruments.


Past newsletters and topics...
Give us your Challenge               Pricing               Get a Quote

 

 What are you looking for? | Let us help | request@minusk.com

What is your Challenge, Need Description or Goal?
First Name*
Last Name*
Email Address*
Telephone*
Company/Organization
Address
City/State-Province  
Country/Postal Code  
Much appreciated, what engine, site or source referred you?
Searched words, other?
Please confirm that you are not a robot






The MK52


NASA Telescope Project

How Our Isolators Work


Spacecraft Vibration Isolation On the Ground




Minus K Technology Inc., Vibration Isolation Systems
460 Hindry Ave., Unit C | Inglewood, CA 90301 | Tel: 310-348-9656 | Contact Us | www.MinusK.com